Facility/Equipment Details
Description
The LCMS-8030 combines world-leading LC performance with advanced mass spectrometry technologies to create a unique system approach to ultra fast mass spectrometry detection.
Utilizing patented technology, the LCMS-8030 features ultrafast multiple reaction monitoring (MRM) transitions, ultrafast mass spectrum measurement speeds, and ultrafast polarity switching to deliver the ultimate in laboratory efficiency. Coupled with high reliability and an array of user-friendly features, the LCMS-8030 is truly the next generation of triple quadrupole mass spectrometry.
Features
Fastest multiple reaction monitoring (MRM) transition times available today (dwell times of 1msec and pause times of 1msec) utilizing patented UFsweeper® technology, which accelerates ions out of the collision cell by forming a pseudo-potential surface.
High-speed multi-analyte detection with 500 MRM transitions in one second
Unsurpassed polarity switching speed (15 msec).
Synchronized survey scan technology, utilizing a high-speed scanning rate of 15,000 u/sec, allows full spectrum scans within a series of MRM measurements
Easy system maintenance—enables greater uptime and usability.
Specifications
Model: LCMS-8030
Mass range: m/z 10 to 2000
Sensitivity: ESI positive: 1 pg reserpine, S/N>3,000: 1 (RMS)
Resolution: R < 0.7 FWHM
Mass stability: 0.05 u / 12 hr
Cross talk: <0.003%
Minimum pause time: 1 msec
Minimum dwell time: 0.8 msec
Scan speed: Max 15,000 u/sec (0.1 u step: 150,000 data points/sec)
Polarity switching time: 15 msec
Interface: ESI
Applicable flow rate: ESI 1 µL/min to 2 mL/min
MRM transition speed: Max 555 channels / sec
DL maximum temperature: 300ºC
Block heater maximum temperature: 500ºC
Utilizing patented technology, the LCMS-8030 features ultrafast multiple reaction monitoring (MRM) transitions, ultrafast mass spectrum measurement speeds, and ultrafast polarity switching to deliver the ultimate in laboratory efficiency. Coupled with high reliability and an array of user-friendly features, the LCMS-8030 is truly the next generation of triple quadrupole mass spectrometry.
Features
Fastest multiple reaction monitoring (MRM) transition times available today (dwell times of 1msec and pause times of 1msec) utilizing patented UFsweeper® technology, which accelerates ions out of the collision cell by forming a pseudo-potential surface.
High-speed multi-analyte detection with 500 MRM transitions in one second
Unsurpassed polarity switching speed (15 msec).
Synchronized survey scan technology, utilizing a high-speed scanning rate of 15,000 u/sec, allows full spectrum scans within a series of MRM measurements
Easy system maintenance—enables greater uptime and usability.
Specifications
Model: LCMS-8030
Mass range: m/z 10 to 2000
Sensitivity: ESI positive: 1 pg reserpine, S/N>3,000: 1 (RMS)
Resolution: R < 0.7 FWHM
Mass stability: 0.05 u / 12 hr
Cross talk: <0.003%
Minimum pause time: 1 msec
Minimum dwell time: 0.8 msec
Scan speed: Max 15,000 u/sec (0.1 u step: 150,000 data points/sec)
Polarity switching time: 15 msec
Interface: ESI
Applicable flow rate: ESI 1 µL/min to 2 mL/min
MRM transition speed: Max 555 channels / sec
DL maximum temperature: 300ºC
Block heater maximum temperature: 500ºC

×