A Magnetorheological Duckbill Valve Micropump for Drug Delivery Applications

Rubayet Hassan, Sevki Cesmeci, Mahmoud Baniasadi, Anthony Palacio, Austin Robbins

Research output: Contribution to journalArticlepeer-review

3 Scopus citations

Abstract

In this study, we propose a duckbill valve microfluidic pump that relies on an electromagnetic actuation mechanism. An FEA/CFD-based approach was adopted for the design of the device due to the coupled electromagnetic–solid–fluid interactions in the device. The simulation methodology was confirmed with the previously published data in the literature to ensure the accuracy of the simulations. The proposed optimum duckbill valve micropump can pump 2.45 µL of fluid during the first 1 s, including both contraction and expansion phases, almost 16.67% more than the basic model. In addition, the model can pump a maximum volume of 0.26 µL of fluid at the end of the contraction phase (at 0.5 s) when the magnetic flux density is at maximum (0.027 T). The use of a duckbill valve in the model also reduces the backflow by almost 7.5 times more than the model without any valve. The proposed device could potentially be used in a broad range of applications, such as an insulin dosing system for Type 1 diabetic patients, artificial organs to transport blood, organ-on-chip applications, and so on.

Original languageEnglish
Article number723
JournalMicromachines
Volume13
Issue number5
DOIs
StatePublished - May 2022

Keywords

  • MRE
  • magnetorheological
  • micropump

Fingerprint

Dive into the research topics of 'A Magnetorheological Duckbill Valve Micropump for Drug Delivery Applications'. Together they form a unique fingerprint.

Cite this