The calibration of a bioelectrical impedance spectroscopy device

Research output: Contribution to book or proceedingConference articlepeer-review

1 Scopus citations

Abstract

A Calibration is one of the key processes to design and implement of Bioelectrical Impedance Spectroscopy. Through the simulation, the specification of hardware can be determined and examined in the ideal condition. After the implementation of the device, the theoretical and measured values are different due to the uncertainties and noises from the environment. In this paper, the calibration procedure will be introduced and discussed for the different sets of electrodes. By adjusting the different values of the active electrodes and applying for the active shields, the measured value will be compared and discussed. Active electrodes and active shields will be a very effective way to isolate and minimize the noises and avoid unwanted stray capacitance values. Further investigations need to be followed to ensure the proposed way to reduce the noises and stray capacitances.

Original languageEnglish
Title of host publicationProceedings of the 2021 Design of Medical Devices Conference, DMD 2021
PublisherAmerican Society of Mechanical Engineers
ISBN (Electronic)9780791884812
DOIs
StatePublished - 2021
Event2021 Design of Medical Devices Conference, DMD 2021 - Virtual, Online
Duration: Apr 12 2021Apr 15 2021

Publication series

NameProceedings of the 2021 Design of Medical Devices Conference, DMD 2021

Conference

Conference2021 Design of Medical Devices Conference, DMD 2021
CityVirtual, Online
Period04/12/2104/15/21

Keywords

  • Active electrode
  • And compensation
  • Bioelectrical impedance spectroscopy
  • Calibration

Fingerprint

Dive into the research topics of 'The calibration of a bioelectrical impedance spectroscopy device'. Together they form a unique fingerprint.

Cite this